When Calibration Passes but QC Fails
A successful calibration does not guarantee that quality-control results will fall within the acceptable range. Calibration and quality control evaluate different parts of the testing process, so it is entirely possible—and important—for calibration to pass while QC fails. When this happens, the QC failure must be investigated before patient results are released. Calibration and QC […]
When Calibration Passes but QC Fails Read More »